Functional Analysis and product breakdown

We can break down the VEGA instrument into two different sub-systems:
- The Interface Optics (Acronym: IOP)
- The Spectrograph (Acronym: SPE)

The main functions of the Interface Optics sub-system are to:
- inject the beams coming from CHARA to VEGA
- adjust the pupil location (see Annex A)
- reduce the size of the beams (from 3/4 inches to 5 mm)
- configure the beams at the entrance of the spectrograph (vertical separation of 10mm)
- provide artificial beams to calibrate and align the instrument
- adjust the optical path difference (the zero OPD plan could be the same as the visible CHARA combiner one)


Due to the important spectral dispersion of the VEGA instrument, it is not necessary to correct the chromatic image shift caused by the atmospheric refraction. In addition, the fact that the main part of the optical path between the telescopes and the instrument is inside vacuum tubes allows to avoid the use of a chromatic OPD compensator (see Annex B).

The main functions of the Spectrograph sub-system are:
- to combine the beams
- to disperse the fringe signal
- to optimize the fringe sampling
- to select the polarization

The following diagram shows the different modules which compose the sub-systems:

subsystem

Sub system decomposition of VEGA

 

 

The following table summarizes the different functions of each module:

SUB-SYSTEM : INTERFACE OPTICS (IOP)

MODULE NAME

ACR.

COMPOSITION

M#

FONCTIONS

Picking Optics

PGO

Set of 4 periscopes

*

Injects the beam coming from CHARA into VEGA

Avoids beam vignetting due to the mounts of the optical elements of the CHARA visible combiner

Beam Compression

BCP

Parabolic and spherical mirrors

 

Reduces the beam size (0.75 inches → 5 mm)

Adjusts the pupil location

Adjusts the ZOPD

Beam Configuration

BCF

Set of 4 flat mirrors

 

Configures the beam (with a beam separation of 10 mm)

Puts the beam in a vertical configuration

Calibration and Alignment Unit

CAU

Artificial sources (spectral and halogen lamps, laser)

Collimator

Mask

*

Produces light for internal alignment and for the co-alignment between the instrument and CHARA

Produces light for the calibration (flat-field, distortion grid, spectral calibration, …)

CHARA/Sources Selector

CSS

Beam splitter

Flat mirror

*

Feeds the spectrograph with the internal sources or with the beams coming from CHARA

SUB-SYSTEM : SPECTROGRAPH (SPE)

MODULE NAME

ACR.

COMPOSITION

M#

FONCTIONS

Focusing Optics

FOC

Parabolic mirror

 

Produces an image at the spectrograph slits level

Slit Module

SLM

Slits

*

Selects a part of the image

Anamorphosis Optics

ANO

2 cylindrical mirrors

 

Optimizes the sampling (ratio 12 between the spectral and the spatial directions)

Collimating Optics

COL

Parabolic mirror

 

Collimates the beams

Polarization Module

POL

Wallaston prism

λ/4 blades

*

Allows to separate the 2 polarization directions

Dispersive Optics

DIS

3 gratings

*

Disperses the light

Selects between 3 different spectral resolutions (1700, 5000, 35000)

Camera Optics

DFC

Parabolic and flat mirrors

 

Produces a dispersed fringes image at the detector level

Neutral Densities

NDN

Neutral densities

*

Attenuates the flux

Detector

DET

Detector

*

Detects the dispersed image

#: if *, motorized or remotely controlled module

Functional analysis of the different subsystems of VEGA.

Implantation


We have studied two possible implantations for the VEGA spectrograph. The first option was to remove the current CHARA visible table and to replace it by a larger one able to host the VEGA interface optic. In fact, this solution is really heavy for CHARA and do not minimize the number of reflexions for VEGA. We present here the preferred solution for the location of VEGA.

VEGA implantation in the CHARA focal laboratory

 

Details of the Zemax solution for the optic interface


The optic interface is separated into the pick-up mirrors installed on the CHARA visible table and a new optical table which will support the interface optics:

Side view of CHARA Visible Table

Top view of CHARA Visible Table
VEGA pick-up systems inside the CHARA visible table

We add motorized pick-up mirrors (periscopic system) in the CHARA visible table in order to pass above the mounts of the optical elements and to send the beams towards the Interface Optic Table.

On the following Figure, we present the current optical design of the interface optic system.

 

ZEMAX view of the Interface Optic Table of VEGA (side view)

 

ZEMAX view of the Interface Optic Table of VEGA (top view)

The new optical interface table will support the Beam Compressors (with compensation of OPD and pupil reimaging function) and the Beam Configuration modules. At the entrance of the table, the beams have a diameter of 0.75” and a horizontal separation of 3”. At the output of the Optic interface table, the four beams are vertically spaced, with a separation of 10mm and the beam size is 5mm. This table will also host the VEGA calibration and alignment sources.


The following figure presents different views of the optical system installed on the interface table:

Optical configuration of the interface optic. Side view (up), front view (bottom)
as seen from the CHARA beams before the pick-up mirrors.